General Measurements and Calculations
7
-
90
LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Mean Pressure Gradient (PG)
To measure Mean PG (Mean Pressure Gradient):
1. From the Doppler Generic Touch Panel, select
Max PG
.
The system displays a trace caliper.
2. To position the caliper at the beginning of the waveform
(V
max
), move the
Trackball
.
3. To fix the start point, press
Set
.
The system displays a second active caliper.
4. To trace the waveform from V
max
to V
d
move the
Trackball
.
5. To complete the measurement, press
Set
.
The system displays TAMAX and Mean PG in the Results
Window.
If Trace Auto is selected, the waveform is automatically traced
after the second caliper is fixed.
If Trace Auto is not selected, manually trace the desired portion
of the waveform.
Automatic
Calculation of
Mean PG
If velocity was previously measured and calculated in Doppler
Mode, Mean PG is calculated from the velocity value when
Mean PG is first selected in the Touch Panel.
The waveform is automatically traced. Velocity and Mean PG
are also displayed on the screen.
If Trace Auto is off, the Auto Trace function is off and traces can
be performed manually.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...