Edge Enhance
Description
Edge Enhance brings out subtle tissue differences and
boundaries by enhancing the gray scale differences
corresponding to the edges of structures. Adjustments to M-
Mode's edge enhancement affects the M-Mode only.
Adjusting
To cycle through settings, adjust
Edge Enhance
on the Touch
Panel.
Values
0-5, with 0 = Off and 5 = High. The current value displays on the
Touch Panel. Values vary by probe, application, and multi
frequency setting. Values are returned to the preset value when
you change: Probe, Exam Category, Exam Calcs, New Patient,
or Multi Frequency.
Benefits
Edge Enhance cleans out the B-Mode image/M-Mode timeline
by subduing some of the gray scale in order to highlight the
vessel wall or organ. This is helpful when you cannot
differentiate between the chambers of the heart.
Affect on other
controls
Edge Enhance operates in realtime only; not in Freeze or CINE.
Rotation (Updown Invert)
Description
You can flip the image up/down.
Adjusting
To flip the image vertically, press Rotation (up/down).
Values
Up/down. Values vary by probe and application. Values are
returned to the preset value when you change: Probe, Exam
Category, Exam Calcs, or New Patient.
Benefits
Beneficial in transvaginal and transrectal scanning.
CAUTION
When reading an rotated image, be careful to observe the
probe orientation to avoid possible confusion over scan
direction or left/right image reversal.
Summary of Contents for LOGIQ 7
Page 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Page 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Page 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Page 983: ...GE Medical Systems ...
Page 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...