8 - 2
Section 8-3 - Monitor
Section 8-3
Monitor
8-3-1
CRT Assy (FRU No. 100)
Purpose: This is a description on how to remove and replace the CRT Assy.
8-3-1-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
8-3-1-2
Needed Manpower
•
2 persons, 5 m travel
8-3-1-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
•
Maneuver Control Console to a suitable position for removing the monitor.
8-3-1-4
Removal procedure
1.) Remove the OP Rear Cover. Refer to section 8-5-10 on page 58.
2.) Unscrew one screw cap located at (1).
Figure 8-1 Removing OP rear Cover and Unscrew screw
(1)