5.4
Changing parameters during the transfer
Introduction
This section describes how to change parameters or add buffer during the electrotransfer
using PWR LARGE SEMIPHOR TRANSPHOR UNIT and PWR SEMIPHOR TRANSPHOR UNIT.
Changing the time or current
Both the time and the current can be changed during the electrotransfer run.
Follow the instructions below to change values during a run.
Action
Step
Select mA or time using the DISPLAY MODE button
.
1
Change the value with the UP and DOWN buttons
. While the value
is being changed, the red LED will blink.
2
Result: About 10 seconds after the values have been changed, the LED will
stop flashing and show the real time run parameters.
50
SEMIPHOR TRANSPHOR UNIT series Operating Instructions 29281777 AA
5 Operation
5.4 Changing parameters during the transfer