Testing instructions for KISS™ and
KISS™ Multilead
Visual Check
Check the application system for external intactness:
• The distributor housing is undamaged.
• The power supply unit with the integrated suction pump is undamaged.
• The trunk cable and the electrode leads are intact. No cracks, kinks or deformations.
• The suction cups have no cracks.
• Surfaces of all silver/silver chloride electrodes are undamaged, i.e., unscratched,
unsoiled.
Remove electrode leads from the electrodes. Remove electrodes from the suction
cups:
• No soiling by electrode gel, hair, etc.
• Moist electrode filter disks must be replaced. Electrode nozzles intact.
For instruments with an extension arm:
Check instrument arm and electrode distributor housing are undamaged and sliding
components lock securely.
System Function Check
1.
Set negative pressure regulator on the electrode distributor to CENTER.
2.
Switch on ECG recorder or connect up power supply unit.
3.
Press suction cups gently onto a smooth even surface and release the cup.
The suction pump starts up and generates the negative pressure.
22745010H
KISS™ / KISS™ Multilead
19
Summary of Contents for KISS
Page 14: ...Functional Description 14 KISS KISS Multilead 22745010H ...
Page 15: ...2 Operating Controls and System Components KISS 22745010H KISS KISS Multilead 15 ...
Page 18: ...Operating Controls and System Components 18 KISS KISS Multilead 22745010H ...
Page 36: ...Testing instructions for KISS and KISS Multilead 36 KISS KISS Multilead 22745010H ...
Page 40: ...Maintenance Cleaning and Troubleshooting KISS KISS Multilead 40 KISS KISS Multilead 22745010H ...
Page 44: ...Spare Parts List KISS KISS Multilead 44 KISS KISS Multilead 22745010H ...
Page 52: ...Schematics 52 KISS KISS Multilead 22745010H ...
Page 53: ......