A.3
Customer-supplied Items
The following items are procured by the customer:
Table A-3: Customer-supplied items
Qty
Description
1
Armoured mains cable to power the product.
1
Grounding cable to connect primary protective earth terminal lug of the product
cabinet to the ground.
>
Liquid-tight flexible metal conduit (½ in. and/or ¾ in.) to provide protection for
the Bushing Adaptors’ cables and Neutral CT cable (working temperature range:
−55 °C to 105 °C
recommended).
Recommended models from Thomas & Betts:
▪
Metal conduit ½ in. liquid-tight, Model ATLA050 (length as required)
▪
Metal conduit ¾ in. liquid-tight, Model ATLA075 (length as required)
>
Liquid-tight metal fittings and gaskets for the Bushing Adaptor.
Recommended model from Thomas & Betts:
▪
Liquid-tight fittings, ½ in. NPT, Model 5332-HT.
¾ in. NPT, Model 5333-HT
▪
Sealing gasket ½ in., Model 5262.
¾ in., Model 5263
>
Plastic conduit with enough length to provide protection for the temperature
sensor cables (optional).
>
Cable trays and cable ties to run the cables.
1
RTV (Room Temperature Vulcanization: silicon-based mastic), 3M Super Silicone
Sealant No. 8663 or equivalent for the MMTS.
1
Heat gun (for heat shrink)
>
General tools (drill, drill bits, screwdrivers, cutter and spanners)