Hydran* M2-X Instruction Manual
Page 94
MA-029
Rev. 1.0,
30-Jul-18
4.4.6
Setting the Readings Parameters
1.
Access
Setup > Readings Setup > Gas Reading Setup
(see Section 3.2.8.1).
Verify and set, as required, the gas reading parameters.
2.
Press
Esc
and access
H2O Reading Setup
(ses Section 3.2.8.2). Verify and set,
as required, the parameters for the moisture and relative humidity average
calculation.
3.
Press
Esc
; verify and set, as required, the
Sens. Temp Avg Per.
parameter
(used for the temperature average calculation of the sensor).
4.4.7
Configuring the Analog Input(s) (If Used)
1.
Access
Service > System Config
>
Current Config
. Verify that the displayed
list corresponds to the Hydran M2-X configuration. If not, access
Service >
System Config
>
Detected Config
; the Hydran M2-X detects the installed
configuration. Press
Save
to reboot the Hydran M2-X.
2.
Access
Setup > I/O Setup > Analog In Setup
(see Section 3.2.9.1).
Note: The following steps must be performed for each analog input interface.
3.
Select the analog input interface to configure; the corresponding parameter
screen is now displayed.
4.
Make sure
Input Range
is set to 4
–
20 mA.
Note: Analog input interfaces are calibrated at the factory for a 0
–
20 mA
range. However, when parameter
Input Range
is set to 4
–
20 mA, the interface
automatically adapts to this range.
5.
In the parameter screen, scroll down the list of parameters and select option
Configuration
. The
Configuration
screen is now displayed.
6.
Select
Identification
and set all parameters as desired.
7.
Press
Esc
to return to the
Configuration
screen and select
Readings Setup
.
Set all parameters as required.
8.
Perform the verification procedure described in Section 4.2.8.
4.4.8
Configuring the Analog Output(s) (If Used)
1.
Access
Service > System Config
>
Current Config
. Verify that the displayed
list corresponds to the Hydran M2-X configuration. If not, access
Service >
System Config
>
Detected Config
; the Hydran M2-X detects the installed
configuration. Press
Save
to reboot the Hydran M2-X.
2.
Access
Setup > I/O Setup > Analog Out Setup
(see Section 3.2.9.2).
Note: The following steps must be performed for each analog output interface.
3.
Select the analog output interface to configure, and then select the option
Configuration
. The
Configuration
screen is now displayed.
4.
Set all parameters as required.