Page 61
Hydran* M2-X Instruction Manual
MA-029
Rev. 1.0,
30-Jul-18
3.6
Service Menu
3.6.1
Sensors Parameters
3.6.1.1
Gas Sensor Param
The
Gas Sensor Param
function gives access to the parameters of the gas
sensor. Each gas sensor has its own set of parameter values; the values are
indicated on the Test Certificate and Data Sheet (for an example, see Figure 3-12)
supplied with the Hydran M2-X. These options do not allow you to modify the
displayed values. To change the values, use
Install New Sensor
(see Section 3.6.2).
The gas sensor parameters are:
•
Gas Serial No.
: Serial number of the sensor (engraved under the sensor
connector). The
same serial number is used for the gas sensor and humidity
sensor.
•
B
,
M
,
N
,
S
and
A1
to
A6
: The ten parameters of the gas sensor in a normalized
format
(integer numbers).
–
The Hydran M2-X computes the gas level in ppm by performing two readings:
the sensor temperature (provided by the thermistor incorporated in the
sensor) and the gas sensor output (gas level). The calculation also requires
these ten parameters (from
B
to
A6
).
–
These ten parameters characterize the behavior of the gas sensor by
modeling its response according to the temperature.
–
The values of these ten parameters are stored in the Hydran M2-X
’s
nonvolatile memory (see Section 2.3.4.3).
•
Gas Param CkSum
: Checksum of the above values. This value is added to the list
to
avoid erroneous values from being entered.
Note: The Hydran M2-X uses the proven technology of the Hydran M2-X sensor. For a
complete description, see Section 2.4.
3.6.1.2
H
2
O Sensor Param
The
H2O Sensor Param
function gives access to the parameters of the moisture
sensor. Each moisture sensor has its own set of parameter values; the values are
indicated on the Test Certificate and Data Sheet (for an example, see Figure 3-12)
supplied with the Hydran M2-X. These options do not allow you to modify the
displayed values. To change the values, use
Install New Sensor
(see Section 3.6.2).
•
H2O SerialNo.
: Serial number of the sensor (engraved under the sensor
connector). The
same serial number is used for the gas sensor and humidity
sensor.
•
H
2
O Param C1
to
C10
: The ten parameters of the moisture sensor in a
normalized
format (integer numbers).