Hydran* 201
i
Installation Guide
Page 26
17996 Rev5, August 2013
Figure 15: Hydran 201
Ti
Assembly
7.
Separate the adaptor and the heating plate.
8.
Protect the adaptor’s threads with the supplied plastic cap, see Figure 7.
Do not remove the protective plastic caps from the sensor
and/or brass adaptor’s threaded extremity until ready to install
on the transformer valve.
9.
Unscrew the sensor to separate it from the brass adapter.
Do not touch the H201 sensor’s membrane for any reason.
10.
Inspect the sensor’s membrane by looking at it. Its surface must be flat,
without cuts or tears.
CAUTION
CAUTION