Page
54
Appendix B — Humilab chilled mirror hygrometer
OUTPUT2
O)TMPC 1)TMPF E)EXIT
Make a selection and continue as with Output 1, or press
E
to
return to the Scale Outputs sub-menu. Press
E
again to return to
the Main menu.
Set balance
type
From the main menu, press 2 to change the type of balance used by
the Electronics Module, adjust the balance interval, or remotely
initiate a balance cycle. The current state will be echoed followed
by a sub-menu:
BALANCE TYPE = PACER
DAY =0
HOUR =12
1) PACER BAL 2) AUTO BAL 3) SET INTERVAL
4) INITIATE BALANCE E) EXIT
Press
1
to set the type of balance to
PACER
(the default) or
2
to set
the type of balance to
AUTO
. Press
3
to set the balance interval. The
following will appear:
DAY=
Type a value for the number of complete days between balance
cycles and press
RETURN
. The following will appear:
HOUR =
Type a value (in 24-hour format — e.g. 2 p.m. = 14) for the number
of fractional days between balance cycles and press
RETURN
. The
interval time will be echoed, and you will be returned to the Set
Balance Type sub-menu.
EXAMPLE
:
Day =0
Balance every 12 hours
Hour = 12
Day = 1
Balance every 24 hours
Hour =0
Day =5
Balance every 5 1/2 days
Hour = 12
At the Set Balance Type sub-menu, press
4
to initiate a balance
cycle. If you do this, the Electronics Module will be returned to