GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 98
7.1.2
Calibration Kits
eXplore Locus SP Calibration Kits contain
the following items:
•
Detector Alignment Plate
1
•
Bebe Grid
2
•
X,Y Ball Centring tool and
adjustment key
3
•
Distance tool
4
•
Wire Phantom
5
•
White nylon screws & washers
6
•
Measured Plastic dowel (value
is indicated on storage tube)
7
•
9" Square
8
•
Precision Dial Indicator
9
•
Dial Indicator Mount
10
Additional tools required include:
•
Shims
•
8mm socket and ratchet
•
T-Allen 5/32"
•
Personal radiation dosimeter
•
custom
Allen keys -
The
Centre the Source
procedure requires an Allen (or hex) key that fits a 9/64" socket, and
has a maximum length of 4";
The
Centre the Stage
procedure requires an Allen (or hex) key that fits a 5/32" socket, and has
a length of 3" and a head of only 0.5";
•
Calibration Worksheets which are found in the Appendix, Chapter 99 (use a printed copy for
recording calibration values and calculations).
1
2
3
8
4
5
6
7
9
10
4"
3"
0.5"
Summary of Contents for Healthcare eXplore Locus SP
Page 2: ...This page is intentionally left blank...
Page 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Page 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Page 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Page 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...