Page 224
7.5.3
Edit Calibration Settings
1.
Access
Calibration Tools
and click the
Calib. Settings
tab.
2.
Click
File
and choose
Read Registry Values
from the drop-down list.
3.
Complete the Centres of Rotation fields as follows:
Centre of Rotation1:
enter the Centre of Rotation as determined during the Centre the Stage
procedure
Centre of Rotation2:
enter ½ of the Centre of Rotation value
Centre of Rotation4:
enter ¼ of the Centre of Rotation value.
7.5.4
Follow Up
Upon completion, back up calibration values.
SAVE CALIBRATION SETTINGS
1.
Capture a screen shot of the calibration values found under the
Calib. Settings
tab in
Calibration Tools
.
2.
Call the file
calibration_values.jpg
3.
Save the document to both C: and D: drives.