Section 7.5 Case/Shielding Removal "Abridged" Calibration
When the case or x-ray shielding is removed from the Locus SP, an "abridged" calibration is
performed which consists of centring the stage, performing verification scans, editing calibration
settings, and follow up. This is also performed whenever the system is moved. The procedures
below use the following applications:
•
System Warm-Up
(C:\Program Files\GE Medical Systems\eXplore Utilities\WarmupApp.exe)
•
Calibration Tools
(C:\Program Files\GE Medical Systems\eXplore Utilities\Calibration.exe)
•
X-Ray and Motor Controller
. The
X-Ray and Motor Controller
(C:\Program Files\GE
Medical Systems\eXplore Utilities\X_Ray_Controller.exe)
•
transgen
program (C:\Program Files\GE Medial Systems\eXplore
Utilities\Calibration\transgen.exe)
7.5.1
Centre the Stage
This procedure is used to centre the stage between the X-Ray Source and the detector.
1.
Open
Windows Explorer
and navigate to D:/. Copy the
calib
folder and rename it
original_calib
.
2.
Turn the unit ON using the key, and press the red
Reset
button.
3.
Warm-up the X-Ray Source using the
System Warm-Up
application as follows:
-
right-click the Warm-Up icon in the task bar and select the
<=10 Minutes to Scan
option
-
once the cycle has finished (approximately 14 minutes), right-click the icon again and
select
Scan Mode
.
The
System Warm-Up
application can be re-initialized by double-clicking on
WarmupApp.exe
in C:\Program Files\GE Medical Systems\eXplore Utilities\
4.
Access the
X-Ray and Motor Controller
program.
5.
Ensure the fields below are set to the following settings:
Enable Left Limit: checked off
Enable Right Limit: checked off
Camera:
Thomson 8M
Bin Mode:
1 x 1
Exposure Time:
3000 ms
kV:
80
µA:
80
Summary of Contents for Healthcare eXplore Locus SP
Page 2: ...This page is intentionally left blank...
Page 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Page 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Page 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Page 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...