GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 191
7.4.6
Edit Calibration Settings
1.
Access
Calibration Tools
and click the
Calib. Settings
tab.
2.
Click
File
and choose
Read Registry Values
from the drop-down list.
3.
Using the Worksheet, enter the Source-Detector distance in the Detector Position field three
times for each binning mode (1x1, 2x2, 4x4),
separating each with a space
(e.g. 249.74 249.74
249.74).
4.
Complete the Centres of Rotation fields as follows:
Centre of Rotation1:
enter the Centre of Rotation as determined during the Centre the Stage
procedure
Centre of Rotation2:
enter ½ of the Centre of Rotation value (note value in Worksheet)
Centre of Rotation4:
enter ¼ of the Centre of Rotation value (note value in Worksheet).
5.
Complete the Central Slices fields as follows:
Central Slice1:
enter the Central Slice value
Central Slice2:
enter ½ the Central Slice value (note value in Worksheet)
Central Slice4:
enter ¼ the Central Slice value (note value in Worksheet)
6.
Complete the Serial Numbers fields as follows:
Tube:
enter the new X-Ray Source’s serial number (e.g. 35106) and note
number in Worksheet
Summary of Contents for Healthcare eXplore Locus SP
Page 2: ...This page is intentionally left blank...
Page 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Page 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Page 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Page 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...