GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 142
7.2.13
Follow Up
Upon completion, back up calibration values.
SAVE CALIBRATION SETTINGS
1.
Capture a screen shot of the calibration values found under the
Calib. Settings
tab in
Calibration Tools
.
2.
Call the file
calibration_values.jpg
3.
Save the document to both C: and D: drives.
4.
Copy D:\calib folder to C: drive.
5.
Copy C:\translations1 to D: drive.
BACK UP REGISTRY
6.
If needed, add the following folder to the C: drive:
Registry Backup
7.
Create a sub-folder under Registry Backup and label the folder with "today’s date" e.g.
yymmdd
8.
Click on the Microsoft
start
button in the bottom right corner of the screen, and choose
Run...
Summary of Contents for Healthcare eXplore Locus SP
Page 2: ...This page is intentionally left blank...
Page 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Page 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Page 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Page 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...