Chapter 12 - - Set Preferences
Page 297
GE H
EALTHCARE
D
IRECTION
5461425-8EN, R
EVISION
B
RIVO
XR118 S
ERVICE
M
ANUAL
Search Range
This parameter is used to change the frequency search range of the grid line
rate configured above. Change this setting if the result of grid line reduction
is not obvious.
64 LP/CM
36LP/CM
66 LP/CM
36LP/CM
68 LP/CM
70LP/CM
70 LP/CM
70LP/CM
72 LP/CM
70LP/CM, 78LP/CM
74 LP/CM
70LP/CM, 78LP/CM
76 LP/CM
78LP/CM
78 LP/CM
78LP/CM
80 LP/CM
70LP/CM, 78LP/CM
Target System Grid Type
Recommend Grid Type for GLR