7.8 Minimum Configuration to Boot/Scan
7.8.1 Minimum Configuration to Scan
1.
Connect the minimum configuration of cables as shown in figures below:
a.
Console (RTU)
b.
USB UI (User Interface) and USB Hub Top
c.
DVI Cable (Digital Visual Interface) from Graphic Card DVI Out to RTV DVI In
d.
Monitor
2.
Connect the mains power cable and mount the pull-out protection, see:
.
3.
Connect the mains power cable to an appropriate mains power outlet and switch ON the circuit
breaker.
4.
Press the ON/OFF Standby button on the control console to boot up the system.
5.
Connect a probe and start an User Program.
Figure 7-25 cable minimum configuration (BT13.5)
Diagnostics/Troubleshooting
Voluson E-Series Service Manual
KTD106657 Revision 2
7-19