Chapter 4 - Functional Checks
4-33
4-5-1
Save Image Settings Only / Save Small Backup (Scan Settings)
The Image/Scan Settings contain:
•
Application Settings
•
User Programs
•
Auto Text
•
3D/4D Programs
•
Scan Assistant
•
Measure Configuration
•
Biopsy Lines
1.) Insert a DVD/CD+R/RW into the drive or connect an external USB device.
2.) Press the
UTILITIES
key on the control panel.
3.) In the “Utilities” menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
4.) On the left side of the screen select BACKUP and then click the
SYSTEM CONFIGURATION
tab.
5.) Click the SAVE button of the “Image Settings Only” resp. “Small Backup (Scan Settings)” group.
6.) Choose the media and click the SAVE button.
7.) Select the NEW FILE... key and enter a file name (without extension).
8.) Click the OK key to start the process. When the saving has been completed, click OK.
Figure 4-23 Save window
Choose media
and
then click [Save]