![GE H48671KK Service Manual Download Page 113](http://html.mh-extra.com/html/ge/h48671kk/h48671kk_service-manual_128609113.webp)
GE H
EALTHCARE
DRAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
Chapter 3 - Setup Instructions
3-47
3-7-4-5
UP-D23MD Printer - Settings
1.) Call up the
‘Printer Preferences’
; operation see:
Section 3-7-4 "Adjustment of Printer Settings"
.
2.) Select the
PAPER
page and set settings as shown in the left image at
3.) Select the
GRAPHICS
page. From the “Color Adjust” pop-up menu select:
a.) Color Balance: Cyan =
0
; Magenta =
0
; Yellow =
0
b.) Gamma Select:
Gamma 1
c.) Color Correction: set
Printer Hardware Color Correction
d.) Lightness: Sharpness =
7
or
8
; Dark =
0
; Gamma =
-12
; Light =
8
4.) Save the adjusted printer settings with APPLY and OK. Close window and exit System Setup.
5.) Assign the Printer to the remote keys
P1
,
P2
,
P3
and/or
P4
; see:
Figure 3-48 Graphics page (Color B Gamma Select)
Figure 3-49 Graphics page (Color Corr Lightness)