GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 8-64
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-25
APS II Assy
NOTE:
APS Power Supply can not be adjusted.
8-2-25-1
Tools
•
Common Phillips screwdrivers
•
Nipper
8-2-25-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-25-3
Removal procedure
1.) Remove the Side Left Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-48
.
2.) Remove the EMI Cover L. Refer to the
8-2-24 "EMI Cover L and R" on page 8-63
.
3.) Remove HDD power cable.
4.) Extract 3 power connector from back plane.
5.) Disconnect the AC input connector.
6.) Remove the Side Right Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-48
.
7.) Remove the EMI Cover R. Refer to the
8-2-24 "EMI Cover L and R" on page 8-63
Figure 8-103 APS II Connectors of backplane
AC Input Connector