Chapter 7 -
Page 7-25
Section 7-8
LED Descriptions
7-8-1
DTRX64II Assy
The DTRX64II Assys are located at the No.1 and No.2 slot in the NEST Assy.
7-8-2
PI128 II Assy (LOGIQ™ P6/P6 Pro)
The PI128 II Assy is located at the front of the NEST Assy.
7-8-3
P3RLY Assy (LOGIQ™ P6/P6 Pro)
The P3RLY Assy is located at the front of the NEST Assy.
7-8-4
SYSCONML Assy
The SYSCONML Assy is located at the No.4 slot of the NEST Assy.
Table 7-10
LED on DTRX64II Assy
LED
Descriptions
Normal State
DS1
FPGA configuration OK status Indicator
Normally Blinking
Table 7-11
LEDs on PI128 II Assy
LED
Descriptions
Normal State
D1
MUX embedded Probe Recognition Indicator
Normally On
D2
Selected probe connector position number bit 0
Depend on live probe position
D3
Selected probe connector position number bit 1
Depend on live probe position
D4
FPGA Configuration OK status indicator
Normally On
Table 7-12
LEDs on P3RLY/P2RLY Assy
LED
Descriptions
Normal State
D1
MUX embedded Probe Recognition Indicator
Normally On
D2
Selected probe connector position number bit 0
Depend on live probe position
D3
Selected probe connector position number bit 1
Depend on live probe position
D4
FPGA Configuration OK status indicator
Normally On
Table 7-13
LEDs on L1SYSCON Assy
LED
Descriptions
Normal State
DS1
CPDI FPGA configuration OK indicator
Normally On
DS2
Frame Start signal. Scan running status indicator
Normally blinking on scanning
DS3
FEBC FPGA configuration OK indicator
Normally On
DS4
DSP operation OK indicator
Normally blinking