8-142
Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-7-7
PCIe Cable Replacement Procedure
8-7-7-1
Tools
Appropriate Phillips screwdriver.
8-7-7-2
Time Required
10 min
8-7-7-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-7-4
PCIe Cable Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower.
3.) Remove the BE module.
4.) Disconnect one side of the PCIe cable from the BIB board and the other side from the BE module.
5.) Remove the PCIe cable.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Figure 8-144 Disconnecting PCIe Cable
PCIe cable connectors
PCIe cable location
Summary of Contents for H45051VN