Top alignment nuts
Nut and bolt set
Studs
Canopy bracket
Housing bracket
Heater door bracket
Radiant heater bracket
Slot
Figure 5-42
Canopy Alignment
6. Remove the steel tube that routes the door cable into the extrusion. Use a 3mm hex key to remove the
screw in the front cable clip. Use a 2.5 mm hex key and 7mm wrench to remove the screw and nut in the
back cable clip.
7. Disconnect the heater door cable. The cable end is accessed through the two holes on the back of the lift
rail. First pry off the hole plug, then rotate the plastic tube so the hex head screw securing the conical cap
faces the rear. Push a small hex key or screwdriver through the small lower hole in the rail, through the
lower hole in the plastic tube inside the extrusion, and back through the rail on the other side of the tube to
hold the tube in place. Insert a 2 mm hex key through the larger hole to loosen the screw that fastens the
conical cap to the plastic tube. Pull the cap out of the rail.
8. Using a 10mm wrench remove the 4 nuts on the housing bracket that secure the entire assembly to the
uprights. Grasp the canopy with both hands and raise it about ½” until the keyhole disengages. Remove
the canopy/heater assembly.
To reinstall the canopy reverse steps. Take care not to pinch the electrical cables.
112
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures