Appendix B: Specifications
B-4
Service Manual
Humidity
Servo Control Accuracy*
± 10% for settings up to 85%; minimum
75% for settings > 85%
Humidity control setting vs
average humidity at 10 cm above
center of bed.
Ramp-up Time
≤
50 minutes
Time to reach 75% RH with a 39°C
control temp from cold start in
25°C, 50% RH room.
Note:
The above item noted with an asterisk (*) indicates essential performance.
Operating Time without Refill
> 12 hours
Operational time at 65% RH
control setting with one filling of
reservoir in 25°C, 50% RH room.
Mechanical
Height
Canopy closed, bed lowered 152 cm
Canopy closed, bed raised 178 cm
Canopy open, bed lowered 209 cm
Canopy open, bed raised 239 cm
Width
66 cm
Depth
114 cm
Weight
149 kg
Mattress Size
48.8 cm X 64.8 cm
Bed Tilt
± 12 degrees continuous tilt
The Maximum Patient Weight
40 kg (88 lbs)
Option Load Limits
Storage Drawer Package
7 kg
Instrument Shelf
9 kg
Monitor Shelf
22 kg
Rotating I.V. Pole
9 kg
Total Each Mounting Rail
22 kg
Uninterruptable Power Supply (UPS) (Not available with 230V devices)
For UPS specifications, refer to the Addendum: UPS for Giraffe and Panda Products (provided with the UPS).
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...