Chapter 7: Repair Procedures
Canopy Removal for Service
Service Manual
7-3
b.
Using a 2.5 mm hex key, remove the four screws that secure the heater housing cover to the
heater door bracket. Remove the heater housing cover. Save the four screws and heater housing
cover for later reinstallation.
c.
Using a 7 mm socket or wrench, remove and discard the four nuts and four star washers across
the north end of the heater door bracket that secure the canopy to the heater door bracket. (Two
of the nuts are accessed though the north access holes in the canopy bracket.) Take note where
cable routing clips are secured, for later reassembly.
Nuts
FIGURE 7-2. Nuts
Access holes
FIGURE 7-3. Access Holes
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...