Chapter 6: Alarms and System Failure Codes
System Failure Codes
Service Manual
6-25
6.2.37 System Failure 63
6.2.38 System Failure 70
6.2.39 System Failure 71
6.2.40 System Failure 81
6.2.41 System Failure 82
Cause(s)
Action(s)
Control board has
incompatible software
version.
Replace control board software with latest version
Cause(s)
Action(s)
Unexpected SBC software
problem
Power cycle the device. If issue persists replace SBC.
Cause(s)
Action(s)
Unexpected SBC software
problem
Power cycle the device. If issue persists replace SBC.
Cause(s)
Action(s)
Unexpected SBC software
problem
Power cycle the device. If issue persists replace SBC.
Cause(s)
Action(s)
Unexpected SBC software
problem
Power cycle the device. If issue persists replace SBC.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...