Chapter 6: Alarms and System Failure Codes
Alarms
Service Manual
6-3
6.1.6 Bed Height Pedals Disabled
6.1.7 Bed Raise Pedal Pressed
6.1.8 Bed Lower Pedal Pressed
Cause or Circumstances
Failure
Action
The Bed Height Pedals Disabled
alarm activates if the bed raise
or lower pedals have been
disabled on either the Service
screen or the Setup screen, and
one of the pedals is pressed.
Bed pedal is disabled in service
Mode.
Bed pedal is disabled in Setup
screen.
Release the pedal.
If the canopy pedal is disabled in
the service screen and you wish to
use it, contact authorized service
personnel to reactivate it on the
service screen.
Cause or Circumstances
Failure
Action
The Raise Pedal Pressed alarm
activates if the pedals are
pressed while turning on the
device.
Pedals are pressed while
turning on the device.
Release the pedal.
Cause or Circumstances
Failure
Action
The Lower Pedal Pressed alarm
activates if the pedals are
pressed while turning on the
device.
Pedals are pressed while
turning on the device.
Release the pedal.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...