8.4.2
Functional Checks
1. Move the light pipe to verify that it moves freely and stays in position.
NOTE:
The light pipe will typically experience some minor deflection (less than 1.3cm) after positioning.
2. Plug the power cord into the power outlet. Confirm that the hour meter display turns on and the
standby/on indicator illuminates in green.
3. Press and release the standby/on switch, and confirm the over-temperature/under-temperature red
indicator flashes briefly and the standby/on indicator changes to blue.
4. Confirm that the fan is operating by momentarily placing your hand close to the exhaust vent when
the device is on to feel the air flow coming out of the device.
5. Confirm that the therapeutic light is on.
WARNING
:
Never look directly at the light. Exposure can cause eye damage.
8.4.3
Light Intensity Check
In order to ensure uniform light distribution and the performance of all the optical components, perform
the following procedure.
1. Turn on the system.
2. Focus the light so that the spot diameter is 35.5 cm (14”) or when the shade is 38 cm (15”) from the
bed surface.
3. Using a calibrated Biliblanket Light Meter II, measure the light at the 5 points indicated on the below
diagram and calculate their average. Confirm that the average is at least 27
µ
W · cm
-2
· nm
-1
. Refer to
symptom S11 in Chapter 10 for troubleshooting.
NOTE:
IEC standards recommend LED module replacement at 25% output reduction, which
corresponds to 27
µ
W · cm
-2
· nm
-1
and occurs at approximately 10000 hours. However, sufficient
irradiance for effective phototherapy continues well after 25% reduction.
1
3
2
4
5
3.5”
3.5”
3.5”
3.5”
53
Chapter 8: Service Maintenance and Checkout
Summary of Contents for Giraffe Blue Spot PT Lite
Page 1: ...Giraffe Blue Spot PT Lite Operation Maintenance and Service Manual GE Healthcare ...
Page 7: ... 7 Part I Operation and Maintenance ...
Page 8: ...8 Part I Operation and Maintenance This page is intentionally left blank ...
Page 16: ...This page is intentionally left blank 16 Chapter 1 Product Overview ...
Page 22: ...This page is intentionally left blank 22 Chapter 2 Product Setup and Operation ...
Page 26: ...This page intentionally left blank 26 Chapter 3 Operator s Maintenance ...
Page 32: ...This page intentionally left blank 32 ...
Page 33: ... 33 Part II Service ...
Page 34: ...34 Part II Service This page is intentionally left blank ...
Page 44: ...44 Important Service Safety Information This page is intentionally left blank ...
Page 50: ...This page is intentionally left blank 50 Chapter 7 Installation ...
Page 55: ...Chapter 9 Calibration No calibration is required for the Giraffe Blue Spot PT Lite 55 ...
Page 56: ...This page is intentionally left blank 56 Chapter 9 Calibration ...
Page 72: ...This page is intentionally left blank 72 Chapter 11 Replacement Procedures ...
Page 88: ...12 5 Wiring Diagram 88 Chapter 12 Service Parts ...
Page 94: ...This page intentionally left blank 94 ...
Page 95: ......