Chapter 4: Operator Troubleshooting
This chapter lists possible symptoms as well as the possible causes and solutions that an operator can
implement. All other possible symptoms and causes require professional diagnostics (Professional service
technicians may refer to Chapter 10).
ID
Symptom Description
Possible Causes
Actions & Solution
OS1
When system is in standby
mode (plugged into power
outlet and standby/
on switch is not pressed
yet), both the standby/on
indicator and hour meter
display are off.
(Normally, when the system
is plugged into a power
outlet, the standby/on
indicator is green and the
hour meter display shows an
hour number.)
Power cord is not properly
connected.
Confirm that the power cord is plugged into
the power outlet and power inlet module. Also
make sure that the power cord is secured by
the retainer clip.
OS2
When the system is plugged
into the power outlet, the
over/under temperature
indicator is illuminated in
steady red
.
The system temperature
is low (under-temperature
condition).
Make sure the system is being used in the
recommended temperature range. Allow
the system to warm up to the ambient
temperature before switching the system
into on mode. (The system fan turns on to
help the system warm up. When the system
has warmed up to the ambient temperature,
the system fan will turn off, the over/under
temperature indicator will turn off, and the
standby/on will illuminate green. This process
may take several minutes.)
27
Summary of Contents for Giraffe Blue Spot PT Lite
Page 1: ...Giraffe Blue Spot PT Lite Operation Maintenance and Service Manual GE Healthcare ...
Page 7: ... 7 Part I Operation and Maintenance ...
Page 8: ...8 Part I Operation and Maintenance This page is intentionally left blank ...
Page 16: ...This page is intentionally left blank 16 Chapter 1 Product Overview ...
Page 22: ...This page is intentionally left blank 22 Chapter 2 Product Setup and Operation ...
Page 26: ...This page intentionally left blank 26 Chapter 3 Operator s Maintenance ...
Page 32: ...This page intentionally left blank 32 ...
Page 33: ... 33 Part II Service ...
Page 34: ...34 Part II Service This page is intentionally left blank ...
Page 44: ...44 Important Service Safety Information This page is intentionally left blank ...
Page 50: ...This page is intentionally left blank 50 Chapter 7 Installation ...
Page 55: ...Chapter 9 Calibration No calibration is required for the Giraffe Blue Spot PT Lite 55 ...
Page 56: ...This page is intentionally left blank 56 Chapter 9 Calibration ...
Page 72: ...This page is intentionally left blank 72 Chapter 11 Replacement Procedures ...
Page 88: ...12 5 Wiring Diagram 88 Chapter 12 Service Parts ...
Page 94: ...This page intentionally left blank 94 ...
Page 95: ......