![GE Ettan IPGphor 3 Operating Instructions Manual Download Page 85](http://html.mh-extra.com/html/ge/ettan-ipgphor-3/ettan-ipgphor-3_operating-instructions-manual_128030085.webp)
Action
Step
The first screen will indicate the protocol, number of steps, and total hours
(excluding rehydration time):
7
S t a r t i n g p r o t o c o l # 1
3 s t e p s 4 : 0 0 H r
If a rehydration period was programmed, the rehydration status screen will
indicate the number of hours elapsed and the actual platform temperature:
R u n n i n g P r o t # 1
R e h y d r a t i o n f o r 1 0 : 0 0 H r
0 : 0 1 H r E l a p s e d
R e h y d r a t e a t 2 0
˚
C
Change rehydration duration or
other parameters
To change rehydration duration or other parameters:
Action
Step
Press EDIT to display the set screen. Pressing EDIT again will access the step
setting screen. Pressing EDIT again will return to the running screen.
1
The first line indicates that the protocol is running.
2
Any changes made while a protocol is running are applied to the current
run only, the stored protocol is not affected. Move the cursor under the digit
you wish to set and change it using the Up and Down arrow keys. To end
this step, set all digits to 0. This screen is automatically replaced by the reg-
ular status screen after a few seconds.
R u n n i n g P r o t # 1
R e h y d r a t e 1 0 : 0 0 H r @ 2 0
˚
C
I E F @ 2 0
˚
C 2 0 0
µ
A / S t r i p
1 2 S t r i p s 3 S t e p s
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC
85
5 Operation
5.3 Perform a run from the control panel
5.3.4 Isoelectric focusing from the control panel