Action
Step
Rinse well with distilled or deionized water.
3
Thoroughly air dry the strip holders/manifold or dry well with a lint-free tissue
prior to use.
4
Wipe the platform and plate electrodes with a lab tissue or paper towel after
each use. Other exterior surfaces can be cleaned with a damp cloth. Never
use abrasive cleanser or solvents.
5
Clean Manifold Light
Clean Ettan IPGphor Manifold Light with water, no warmer than 40°C, and with IPGphor
Strip Holder Cleaning Solution.
CAUTION
Do not use any organic solvents during cleaning.
Clean electrodes
NOTICE
Avoid scratching the electrode surfaces.
Wash the electrode assemblies with the IPGphor Strip Holder Cleaning Solution. Rinse
thoroughly with water then deionized water and allow to air dry.
Cleaning before planned
maintenance/service
To ensure the protection and safety of service personnel, all equipment and work areas
must be clean and free of any hazardous contaminants before a Service Engineer starts
maintenance work.
Please complete the checklist in the On Site Service Health and Safety Declaration Form
or the Health and Safety Declaration Form for Product Return or Servicing, depending on
whether the instrument is going to be serviced on site or returned for service, respectively.