User Manual
Druck DPI612
Page 43 of 105
[EN] English – 109M4017 Revision -
2.2.3
Vacuum or Pressure Operation DPI 612 pFlex (20 bar)
1
2
3
4
5
Figure 2-5 Vacuum or Pressure operation
2.2.1
Release Pressure DPI 612 pFlex (20 bar)
Open the pressure release valve counter clockwise (1 turn)
Table 2-2
Table 2-3
Step Procedure (Vacuum)
Step Procedure (Pressure)
1
Set to vacuum operation
(-).
1
Set to Pressure operation
(+).
2
To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster to the
middle of its range of
operation.
2
To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster to the
middle of its range of
operation.
3
Seal the system.
3
Seal the system.
4
Use the pump to set the
maximum vacuum or set
the vacuum you want to
adjust.
4
Use the pump to set the
maximum pressure or set
the pressure you want to
adjust.
5
Adjust the vacuum
(+ decrease; - increase).
5
Adjust the pressure
(+ decrease; - increase).