Changing the Units
© 2017 General Electric Company – All Rights Reserved.
ADTS 405 R/F Mk2 Instruction Manual-English | 31
4.2.1 Change to Single, Dual or Triple Display
a.
Press
SETUP
.
b.
Select [MORE].
c.
Select [DISPLAYS/OPTIONS].
d.
Select [DISPLAY TYPE].
e.
Using [NEXT] or [PREV].
f.
Select either SINGLE, DUAL or TRIPLE, ENCODER (more display types with options
fitted).
g.
Press [SAVE] to accept.
h.
Press
CLEAR/QUIT
repeatedly to return to user display.
4.2.2 Changing the Display for Pt Only
a.
Press
SETUP
.
b.
Select [MORE].
c.
Select [Control], [Ps Pt Dual].
d.
Select [Pt Only Mode].
e.
Press
CLEAR/QUIT
repeatedly to return to user display.
4.3 Changing the Units
In Full Setup Mode any set of aeronautical or pressure units can be selected.
4.3.1 Changing the Measurement Units
a.
Press the
SETUP
key.
b.
Select [UNITS].
c.
Use [NEXT] and [PREV] keys to select the required units.
d.
Select [SAVE] to accept.
e.
Press
CLEAR/QUIT
repeatedly to return to the main pressure display.
Note:
(1) In Minimum Setup Mode only the default aeronautical and pressure units can
be selected.
Note:
(2) If Setup Mode is switched off, units cannot be changed.
4.3.2 Changing the Unit Type
a.
Press the
SETUP
key.
b.
Select [UNITS TYPE].
c.
Select either [AERO] or [PRESS] to select the unit type.
d.
Select [SAVE] to accept.
e.
Press
CLEAR/QUIT
repeatedly to return to the main pressure display.
Note:
(1) The default pressure and aeronautical units are defined in the CONFIG mode.