Revision C
DINAMAP
®
ProCare Monitor Service Manual
4-9
2009381-001
Maintenance:
Parameter Level Functional Testing
Replacement batteries can be obtained from GE Medical Systems
Information
Technologies
.
NOTE:
:
The replacement part number of the battery is 633178CR. Use of
nonspecified batteries can cause internal damage to the Monitor or
cause patient/user safety issues.
Fuses
The Monitor contains four fuses. The fuses are auto-resettable and mounted
within the Monitor. The fuses protect the low voltage DC input, the battery, the
remote alarm output, and the +5 V output on the host port connector.
Parameter Level Functional Testing
After the initial configuration is complete, perform functional testing of each of
the parameters, using the accessories supplied with the ProCare Monitor.
Refer to the operation manual for more detailed parameter-specific instructions.
BP
Perform a blood pressure by connecting the supplied hose and cuff together,
then attaching to the front of the ProCare Monitor. Press the
Start
button on the
front to begin the BP cycle.
Temperature
Connect the supplied temperature probe to the corresponding connection. A
predictive temperature will begin once the probe is removed from its holster.
Place the probe in the holster after completion of the Temp cycle.
Nellcor SpO
2
Technology
The SpO
2
sensor is an assembly consisting of two parts: the DS-100A sensor and
the DOC-10 extender cable. Connect the cables prior to attaching to the Monitor.
A SpO
2
reading will be displayed within moments of attaching the sensor to
either a Nellcor simulator or to your finger.
Masimo SpO
2
Technology
The SpO
2
sensor is an assembly consisting of two parts: the LNOP NR-195 sensor
and the PC08 extender cable. Connect the cables prior to attaching to the
Monitor. A SpO
2
reading will be displayed within moments of attaching the
sensor to either a Nellcor simulator or to your finger.
Summary of Contents for DINAMAP ProCare
Page 2: ......
Page 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Page 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Page 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Page 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Page 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Page 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 133: ......