Revision C
DINAMAP
®
ProCare Monitor Service Manual
3-11
2009381-001
Installation:
Configuring Your ProCare Monitor
Service Mode
Service mode is entered by holding the
Cycle
button while powering on with the
On/Off
button. You can press the
Cycle
button to advance through the available
service mode pages.
NOTES
Only transducer calibration pages are available until calibration is valid.
Calibration and other service mode setting changes will not be retained
unless the “Save Settings” operation is executed (on the final Service Mode
options page).
After all options pages have been displayed, the display will return to the 1
st
service mode page (initial calibration page). To save service mode settings the
Service Mode option pages are as follows:
Displayed on Monitor
Function
0 (in “min” window)
Refer to calibration section for functions.
1 (in “min” window)
2 (in “min” window)
3 (in “min” window)
NIBP Algorithm Type loaded
(Displayed in Systolic display window)
1: Classic DINAMAP NIBP
2: Auscultatory NIBP
(This setting cannot be changed.)
4 (in “min” window)
SpO
2
Type loaded
(Displayed in SpO
2
% display window)
0: No SpO2
1: NellCor
2: Masimo
(Warning: Incorrect setting will cause fatal 930 alarm
during operation.)
5 (in “min” window)
Temperature loaded
(Displayed in Temperature display window)
0: No Temp
1: Turbo Temp
Summary of Contents for DINAMAP ProCare
Page 2: ......
Page 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Page 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Page 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Page 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Page 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Page 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 133: ......