4-16
ProCare
Patient Monitor Service Manual
Revision A
2009381-001
Calibration & Maintenance:
4.6 ProCare Patient Monitor Parameter Tests
5. Record and verify systolic, diastolic, map and heart rate from the monitor
display
6. Press
CYCLE
button until
is displayed on the HISTORY LED to
initiate a determination in STAT mode.
7. Record and verify systolic, diastolic, map and heart rate from the monitor
display
8. Press
Start/Stop
button, end STAT mode.
4.6.10 NIBP Overpressure Verification
1. Restrict airflow through cuff hose port.
2. Press "
Inflate/Stop
" to begin NIBP determination
3. Verify that
is displayed on the
SYSTOLIC
display and an audible
alarm sounds
4. Remove the air restriction
5. Press "
Inflate/Stop
" and verify that the pump does not start
6. Press the "
Silence
" button
7. Press the "
Silence
" button again
8. Verify the alarm condition is cleared from the
SYSTOLIC
display
4.6.11 Temperature (Perform if equipped with Temp module)
The Temperature Simulator for the Alaris System is available from Alaris
Medical Systems, Inc. (619) 458-7000.
1. Disconnect the temp probe from the Monitor.
2. Connect the probe simulator to the Monitor.
4. Set the probe simulator to 80.2° F.
5. Record and verify the reading in the
TEMP
display is 80.2° F ±0.2° F.
6. Set the probe simulator to 102° F.
7. Record and verify the reading in the
TEMP
display is 102° F ±0.2° F.
8. Set the probe simulator to B.P. and verify reading is 106.0° F ± 0.2° F.
Press broken probe button down and verify the Monitor displays
.
Summary of Contents for DINAMAP ProCare 100
Page 1: ......
Page 4: ...DINAMAP ProCare Monitor Service Manual 2009381 001 Revision A ...
Page 9: ...ProCare Patient Monitor Service Manual Revision A 2009381 001 Table of Contents ...
Page 10: ...Revision A ProCare Patient Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 11: ...1 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 ...
Page 20: ...Revsion A ProCare Partient Monitor Service Manual 2 1 2009381 001 2 Product Description ...
Page 21: ...2 2 ProCare Partient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 36: ...Revision A ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Principles of Operation ...
Page 37: ...3 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 46: ...Revision A ProCare Patient Monitor Service Manual 4 1 2009381 001 4 Calibration Maintenance ...
Page 47: ...4 2 ProCare Patient Monitor Service Manual Revision A 2009381 001 For your notes ...
Page 83: ...PNEUMATIC TUBING ROUTING Exploded Assembly Drawing ProCare Monitor front case page 5 3 4 or ...
Page 96: ...2009381 001 A ...