Chapter 1. Calibration
8
DigitalFlow™ GS868 Service Manual
1.5.5 20 mA Option = [F4]
1.
Connect the 20 mA current source to the currently selected analog input, as shown in Figure 3 on page 6.
2.
Press
[F1]
to store the current 20 mA value or press
[F2]
to cancel the entry. In either case, the
ANALOG INPUT
prompt
reappears.
3.
Go to
Procedure Options
below.
Procedure Options
You have completed calibration of the analog output. Do one of the following:
•
To calibrate another input, return to Step 3 in
Accessing the Calibration Menu
on the previous page.
•
To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
twice.