DigitalFlow™ GS868 Service Manual
23
Chapter 2. Error Codes and Screen Messages
End Time must exceed
Start Time
by 5 min.
This message appears when in the
LOG
menu. Enter an end time that is at
least five minutes later than the start time.
Header invalid
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Log Active, END only
hit any key
When in the
LOG
menu, this message indicates that the log is still
compiling data. Only the End Time may be edited.
Log DONE, to inspect
hit any key
When in the
LOG
menu, this message indicates that the log is complete. Hit
any key to display the log.
Log hasn’t started! hit
key
When in the
LOG
menu, this message indicates that the log has not yet
started.
Outside limits, value
rejected.
When in the
CAL
menu, this message indicates that the calibration of the
analog output is invalid. Hit any key to clear the message, and the GS868
will default to the last valid calibration. The message will also appear if
there is no ammeter connected to the analog output during calibration.
Overflow
The display value overflowed. Reduce the number of decimal places or
change the units.
??P<L Enter L again.
The entered path length (P) is less than the axial dimension (L). Enter a new
value for L.
range is X.XX to
X.XX
When in the
User Program
, this message indicates that the entered number
is unacceptable. Hit any key and enter a number within the allowable range.
Review calibration
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Review parameters
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Staring time must
exceed current time
This message appears when in the
LOG
menu
STD
option. Enter a start time
that is later than the current time.
Write error.
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Table 2: Screen Messages (cont.)
Message
Meaning
Summary of Contents for DigitalFlow GF868
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii DigitalFlow GS868 Service Manual ...
Page 26: ...Chapter 1 Calibration DigitalFlow GS868 Service Manual 16 ...
Page 40: ...Chapter 3 Diagnostics 30 DigitalFlow GS868 Service Manual no content intended for this page ...
Page 68: ...Warranty 58 DigitalFlow GS868 Service Manual no content intended for this page ...
Page 72: ......
Page 73: ......