Chapter 1. Calibration
8
DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
1.5 Calibrating the RTD Inputs
Calibrating an RTD option card involves a slightly different procedure than for other analog input cards. However, you
access the card in the same manner.
1.5.1 Accessing the Calibration Menu
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate the desired slot. (The option bar will include a slot listing for each installed option card.)
3.
Press
[F1]
to select Input A, or
[F2]
to select Input B.
4.
Press
[F1]-[F2]
to select the reference point to be calibrated.
Do one of the following:
•
If you selected
[F1]
(LOW),
proceed to step 5.
•
If you selected
[F2] (HIGH),
proceed to step 8.
1.5.2 LOW Option = [F1]
5.
Enter the low reference value and press the
[ENT]
key.
6.
Press
[F1]
to
STORE
the low reference value or
[F2]
to
ABORT
the calibration. In either case, the screen returns to the
ANALOG INPUT
prompt.
7.
Press
[F2]
to proceed to the next section.
1.5.3 HIGH Option = [F2]
8.
Enter the high reference value and press the
[ENT]
key.
9.
Press
[F1]
to
STORE
the current high reference value or press
[F2]
to
ABORT
the entry. In either case, the
ANALOG
INPUT
prompt will reappear.
10.
Proceed to
Procedure Options
below.
Procedure Options
You have completed calibration of the RTD input. Do one of the following:
•
To calibrate another input, press
[EXIT].
•
To calibrate additional slots, press
[EXIT]
and then
[F4] (EXIT)
and proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
twice.