November 2005
6-2
Technical Specifications
Optional Onboard
Display with User
Interface
The optional onboard display with user interface uses a matrix
configurator for:
•
range changes
•
unit of measure selection
•
current loop adjustment
•
error diagnostics
•
current value selection for fault conditions
•
and entering a pressure constant for ppm
v
.
EMI/RFI
Performance Criterion A:
1.
Conducted Emission Test as per CISPR 11 Class A, 2004
2.
Radiated Emission Test as per CISPR 11 Class A, 2004
3.
Radiated Susceptibility Test as per IEC 61000-4-3, 2002
4.
Electrostatic Discharge Test as per IEC 61000-4-2, 2001
5.
Electrical Fast Transient Test as per IEC 61000-4-4, 2004
6.
High Energy Surge Immunity Test as per IEC 61000-4-5, 2001
7.
Power Frequency Magnetic Field Test as per IEC 61000-4-8, 2001
EMC
IEC 61326, Industrial Locations
Optional Certifications/
Approvals
FM IS Cl. I, II, III, Div. 1, Grps. A-G, T5
FM XP-IS Cl. I, Div. 1, Grps. A-D, T5
FM NI Cl. I, Div. 2, Grps. A-D, T4A
DIP Cl. II, III, Div. 1, Grps. E-G, T5
ATEX II 3G EEx nA IIC T4