![GE Datex-Ohmeda S/5 Technical Reference Manual Download Page 212](http://html.mh-extra.com/html/ge/datex-ohmeda-s-5/datex-ohmeda-s-5_technical-reference-manual_128582212.webp)
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
76
Document no
.
M1137266-02
6.2 Error History
6.3 Event History
Last Errors:
The section shows the last monitor
errors and the time of their occurence.
Error counters:
The section contains counters for
each different (detected) monitor error. The time of
occurrence of the last error is shown beside each
counter.
Last log reset:
The date and time of the last Error/
Event log reset.
Possible errors:
Fast cold start
indicates the number of erroneous
cold starts with power off time less than 20 seconds.
The reason can be either a failing lead acid battery
or a software problem that was solved by the
hardware watchdog circuitry located in the
monitor’s power supply unit.
GSP watch-dog timeout
indicates an erroneous
restart controlled by the main CPU board and
caused by the display controller board. The restart is
listed as a fast cold start.
UPI watch-dog timeout
indicates an erroneous restart controlled by the main CPU board and caused by the
UPI or UPINET board. The restart is listed as a fast cold start.
Last events:
The section shows the last events and
the time of their occurrence.
Event counters:
The section contains counters for
each different (detected) event. The time of
occurrence of the last event is shown beside each
counter.
Last log reset:
The date and time of the last Error/
Event log reset.
Possible events:
Cold start
is a start-up with power off time more
than 15 minutes. The trend memory is cleared and
monitoring starts with the user default settings.
Warm start
is a start-up with power off time less
than 15 minutes. The trend information and
possible temporary settings are still available.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...