Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
20
Document no
.
M1125633-06
3.
Check that the message ‘Cable off’ is shown in the digit field and that ‘Cable’ on the
service menu states
OFF
.
Plug the NMT Sensor Cable with the NMT ElectroSensor into the front panel connector
NMT. Check that the message in the digit field changes to ‘Measurement OFF’ and ‘Cable’
in the service menu states
EMG
and
ELECTR. OFF
.
4.
Place a 3 k
Ω
resistor between the ElectroSensor’s stimulus electrode leads (brown and
white). Perform
Start Curr. Test.
Check that the test was successful, i.e. the ‘Current test (mA):’ in the menu states
30 OK, 50
OK
and
70 OK
.
Connect the ElectroSensor leads to the NMT simulator. Set the switch on the simulator to
FADE OFF and turn the knob to MAX. Check that ‘Cable’ in the service menu states now
only
EMG
.
5.
Start the NMT measurement (TOF) by pressing the
START-UP
key on the module.
Check that the found supramaximal current is less than 70 mA, i.e. the ‘Current set’ value
in the service menu is less than 700.
6.
Check that the digit field TOF% value is within 95-105, Count is 4 and T1% is within
95-105.
7.
Check that the Noise value in the service menu remains under 100.
"
2.2.16 Nellcor Compatible Saturation Module, E-NSATX
1.
Enter the
N/O-SAT
service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the NSAT module’s ROM memory has passed the
internal memory test, i.e. ROM shows OK.
Check that all three error indicators, Preamp Error, QUART Error and I/O Error state NO.
2.
Check that the SpO
2
probe related status information in the menu is correct. Only the ‘No
Probe’ should be active (1), when no probe is connected.
3.
Connect a Nellcor SpO
2
finger probe to the module. Check that the message ‘No Probe’
changes to ‘Check Probe.’
4.
Attach the SpO
2
probe on your finger. Check that a reading of 95-100 and a proper SpO
2
waveform appear.
"
2.2.17 Masimo Compatible Saturation Module, E-MASIMO
1.
Enter the
N/O-SAT
service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check also that the module’s ROM memory has passed the
internal memory test, i.e. the ROM shows OK.
Check that all three error indicators, Preamp Error, QUART Error and I/O Error state NO.
2.
Check that the message ‘No probe’ is displayed on the screen when no probe is
connected.
3.
Connect a suitable SpO
2
finger probe to the module. Check that the message ‘Check
probe’ is displayed on the screen within 30 seconds.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...