Planned maintenance check list
17
Document no. M1125633-06
2.
Check the
BIS
and
Sensor Check
membrane keys of the module. Stay in the module
view and press each key for at least one second and check that the key being pressed is
identified, i.e. the corresponding 'PUSHED' text appears in the service menu.
3.
Check that ‘Messages from BE’ are increasing steadily.
Enter the
Sensor
menu
.
Check that
- sensor type indicates 'No Sensor' (i.e. no sensor is identified)
- no sensor is identified
- mains frequency is set correctly
- check that ‘BE powerup test’, ‘DSC selftest Ch1’ and ‘DSC selftest Ch2’ all show 'PASSED'
(if not, go to
BIS Setup
menu, perform DSC Test and check the results again)
4.
Connect the BIS simulator to the PIC+ cable. See that ‘Checking sensor’ text and image
appear in the waveform numeric field. Wait for a while and check that all sensors show
'PASS'. Check that the ‘Sensor type’ shows correct information.
5.
Check that the ‘BIS’, ‘SQI’ and ‘SR’ values are between 0...1000, and the ‘EMG’ value
between 0...10000.
NOTE: If Sensor Simulator 900508 is used, the values can be out of the given range.
6.
Enter the
Module
menu.
Check that no BIS engine errors appear.
Perform sensor check by pressing
Check Sensor
and see that the check is passed.
During the sensor check the Impedance meas. should show 'CYCLIC'. During operation
the Impedance meas. should show 'COMBINED' or 'OFF' (depending on the user setting in
BIS Setup
menu).
"
2.2.11 Entropy Module, E-ENTROPY
1.
Enter the
Entropy
service menu.
Check that the Entropy sensor cable is connected to the module.
Check that the Time-outs, Bad checksums and Bad c-s by mod values in the module view
are not increasing faster than by 5 per second. Check that the memories of the module
have passed the internal memory test, i.e. RAM, ROM and EEPROM all state OK.
2.
Check the
Entropy
and
Check Sensor
membrane keys of the module. Stay in the
service menu and press each key for at least one second and check that the key is
identified, i.e. highlighted when pressed.
3.
Connect the Entropy simulator (N-ES) to the cable. See that the ‘Checking sensor’ text and
image appear in the waveform numeric field. Wait for a while and check that all sensors
show “PASS”.
4.
Check that the EntrEEG waveform and RE and SE values appear on the monitor screen.
"
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...