Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
12
Document no
.
M1125633-06
2.2.5 Extension Frame, F-EXT4
1.
If the F-EXT4 contains a fan, check that the fan is running.
2.
Check that the modules in the F-EXT4 are recognized, i.e. the required parameter
information is shown on the monitor screen.
Disconnect the Extension Module, E-EXT shortly, then reconnect the module back into the
F-CU8.
Check that the modules in the F-EXT4 are still recognized.
"
2.2.6 Compact Airway Module, E-CXXXXX
1.
Check that the fan is running.
2.
If the module contains membrane keys on the front panel, press each of the keys for at
least one second and check that they are identified.
3.
Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter
the Compact Airway Module
General
service menu.
Check that the module configuration displayed corresponds with the Compact Airway
Module type being used.
4.
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second.
5.
Enter the
Gases
service menu: Check that the ‘Ambient’ value displayed corresponds with
the current ambient pressure (±20 mmHg).
6.
Check that the flow measurement offset, i.e. the sample ‘Zero’ value displayed is within
±10 ml/min.
7.
Perform a sampling system leak test.
8.
Check the flow rates. Adjust the sampling flow, if necessary.
9.
Perform a gas calibration:
NOTE: For maximum accuracy, a warm-up time of 30 minutes is recommended.
NOTE: For correct measurement values, modules need different amounts of oxygen in the
calibration. If you do not use the recommended calibrating gases, the calibration does not
succeed.
"
Anesthesia Agent
Compact Airway Modules with the option
10. Enter the
Gases
service menu.
Feed calibrating gas (order code 755583) continuously for at least 30 seconds and check
that the ‘ID’ in the service menu shows ‘DES’ and that the value for ‘ID unrel.’ is lower than
75.
"
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...