Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
40
Document no
.
M1144951-004
Figure 14
Connection cables and LED indicators
WARNING
The Compact Monitor, interfacing modules and interfaced devices must be
situated in the same patient environment (as defined in IEC 60601-1-1).
WARNING
Connecting electrical equipment together or using the same extension cord
for more than one device may cause their leakage currents to exceed the
limits specified in relevant safety standards. Always make sure that the
combination complies with the international safety standard IEC 60601-1-1
for medical electrical systems and with the requirements of local
authorities.
WARNING
The manufacturer guarantees a reliable functioning of the devices with
tested software versions only. Always refer to the Installation guide
accompanying the DIS module and verify the compatibility before use.
4.2.3 Mounting
The DIS module can be mounted on the side panel of the external device. Also IV pole
placement is possible.
NOTE: As the Device Interfacing Solution works only with the device specified in the label of the
interfacing module, it is recommended that the interfacing module always travels along with
the external device.
For mounting accessories, please refer to the “Supplies and Accessories” catalog. See the
following figure for an example of a device interfacing.
(1)
label specifying the external device
(2) LED
indicators
(3)
black bus cable from another
interfacing module, if needed
(4)
grey device specific cable to the
communication port of the external
device
(5)
black bus cable to the monitor’s DIS
connector (or to another interfacing
module)