Display Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
23
Document no. M1144956-003
3.2.5 Functional inspection
9.
Stand-by LED
Connect the power cord.
Check that the standby LED starts blinking or is on continuously, if the batteries do not
need charging.
"
10. Start-up
Switch the monitor on.
Check that the monitor starts up properly, i.e. the alarm LEDs turn blank, the start-up
sound is heard from the loudspeaker and the normal monitoring screen appears.
No error messages should appear on the screen.
If the monitor contains the recorder unit, two lines of start-up information should be
recorded.
"
11. Monitor screen configuration
Configure the monitor screen so that all the needed parameter information is shown:
Monitor Setup
Screen 1 Setup
Waveform Fields
Digit Fields
Check that the selected parameter information is shown on the screen.
NOTE: InvBP waveforms are not shown without a patient simulator. Do no connect the
simulator to the monitor at this stage.
"
12. Mark Event or Take Snapshot
Push the menu keys
Help
and
Mark Event
, or
Take Snapshot.
Check that the picture on the LCD display screen is straight, clear and stable.
"
13. Time and date
Check that the time and date are correct, adjust them if necessary.
Monitor Setup
Set Time and Date
NOTE: The selection SET TIME AND DATE cannot be accessed if a case is started, or a patient
has been admitted.
NOTE: If the clock shows time 0:00 continuously (at successive start-ups), the SRAM/
TIMEKEEPER battery needs to be replaced.
The FACTORY RESET should be performed after the battery replacement,
"