4-14
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Maintenance:
Lithium-Ion Battery
Identifying Battery Capacity Using the Cadex SMart Two+ Charger
When you select the
Target Capacity
switch on the charger, the
charger compares the battery’s performance to a 60%, 70%, or 80% target
capacity set on the battery charger. If the battery fails to meet the target
performance, the battery charger will prompt you to condition the
battery. If after the condition cycle has been completed, the battery does
NOT meet the target performance, then the battery charger will
illuminate a “fail” light.
Identifying Battery Status Using the Battery Status Menu
This menu option opens a menu and information window that provides
current battery status information.
1. Select
MORE MENUS
from the Main Menu.
2. Select
BATTERY STATUS
to display the Battery Status Menu and
information window.
BATTERY STATUS
Menu and Information Window
If a battery is not present, a
NO BATT
message is displayed in
this column.
If the battery is
NO COMM
(communication with this battery
has failed), “unknown” is displayed for all rows except the
SLOT
STATUS
row.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......