Equipment Overview:
Technical Specifications
Analog Output
Defibrillator Synchronization Pulse
ECG:
Gain:
DC offset:
Noise:
Frequency response:
Time delay:
1 V/mV ±10%
±100 mV (max)
<5 mVp-p (0-300 Hz)
In the
section, refer to the
40 ms monitoring filter, 35 ms diagnostic filter
Blood pressure:
Gain:
DC offset:
Noise:
Frequency response:
Time delay:
10 mV/mmHg ±2%
±20 mV (max)
<5 mVp-p (0-300 Hz)
dc to 50 Hz +2/–0 Hz
40 Hz filter, 37 ms
Marker out:
Time delay:
Amplitude (selectable in Service
menu):
+5 V selection:
+12 V selection:
Pulse width:
Output impedance:
Current limit:
35 ms (maximum), R-wave peak to leading edge of
pulse.
3.5 V (min) at 1 mA sourcing; 0.5 V (max) at 5 mA
sinking.
11.0 V (min) at 1 mA sourcing; 0.75 V (max) at 5 mA
sinking.
10 ms ±10% or 100 ms ±10% (selectable in Service
menu).
50
Ω
nominal
15 mA nominal, both sourcing and sinking.
Marker in:
Input threshold:
Input hysteresis:
Maximum input voltage:
Input impedance:
Pulse width:
V
IH
= ±2.5 V (min); V
IL
= ±1.5 V (max)
650 mV typical
±30 V (with respect to ground on pin 3)
10 k
Ω
(min) for -25 V < VIN < 25 V
1.0 ms (min), VIN > 2.5 V
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......