Troubleshooting:
Wireless LAN Troubleshooting
Wireless LAN Troubleshooting
Problems may occur with network communications while the monitor is
communicating over Wireless LAN. Following is a list of possible
solutions when troubleshooting Wireless LAN.
If the monitor can not be viewed at the central station or can not view
other monitors, it may not be communicating with the internal RF card.
1. Check the status of the Wireless LAN communication:
a. Select
REVISION AND ID
option from the Monitor Setup Menu.
b. There are three pages of software revision information. Select
NEXT
to display the second page of information.
c. Select
NEXT
twice to display the Hardware ID Display
information window.
d. Verify the Wireless LAN ID shows the proper country, status,
revision and date code. For example: US 02 V4.63 00022.
2. If there is no Wireless LAN ID, the card may be disabled or
inoperative. Verify the card is enabled:
a. Activate the Boot Code by holding down
NBP GO/STOP
and
ZERO ALL
.
b. Press and release the
Trim Knob
control.
c. Keep holding
NBP GO/STOP
and
ZERO ALL
until the Boot
Code information appears on the display.
d. Select
SERVICE MENU >
SET CONFIGURATION
>
CONFIGURE WIRELESS LAN
>
ENABLE/DISABLE
WIRELESS LAN.
3. The monitor is configured with international roaming. The country
abbreviation code (US 02 V4.63 00022) must match the country the
monitor is in. If the code does not match:
a. Check to make sure the correct Access Point is installed.
b. Refer to the Wireless LAN (Symbol Access Point) Installation
and Service Manual for details on proper Access Point
installation.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......