5-22
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Troubleshooting:
Data Acquisition Tests
Respiration Functions
Connect the Multifunction Microsimulator, pn. MARQII, and
appropriate patient cables to the ECG/RESP side panel connector on the
monitor.
1. Adjust the patient simulator to output a respiration waveform using
the following settings:
Rate BPM – 30
Baseline Impedance Ohms – 750 (1000 Ohms MARQII),
∆
R Ohms – 2.0.
2. Enable the respiration function of the monitor.
a. Select
MONITOR SETUP
from the main menu display on the
monitor,
b. Select
PARAMETERS ON/OFF
from the monitor setup menu.
c. Turn and push the
Trim Knob
control to scroll to and select
RR
in the parameters on/off pop-up window.
d. Toggle and select
ON
in the RR line of the parameters on/off pop-
up window.
3. Verify the following:
Respiration rate is displayed and accurate.
Respiration waveform is displayed and noise-free.
Markers appear in the displayed respiration waveform (see
figure below). These indicate the points at which the monitor
senses inspiration and expiration for determination of the
respiration rate.
582A
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......